Ion Chip capacities for Ion AmpliSeq™ DNA libraries sequenced at equal depths

The number of combined libraries that can be accommodated in a single sequencing run depends on the chip, the balance of barcoded library concentration, and the coverage required.

For a given chip, as the number of amplicons increases, the number of libraries that can be accommodated per sequencing run decreases. This relationship is shown in the following table. The numbers in the following table serve as a guide for approximate capacities. As the number of libraries per chip increases, it becomes more difficult to balance the reads between libraries. In addition, libraries from FFPE tissue tend to produce more variable results. We suggest combining fewer libraries initially and determining real limits empirically.

Chip

Ion 314™[1]

Ion 316™

Ion 510™

Ion 318™

Ion 520™

Ion 530™

Ion PI™

Ion 540™

Ion 550™

Average depth

150X

2500X

150X

2500X

150X

2500X

150X

2500X

150X

2500X

150X

2500X

Expected coverage

95%
>30X

95%
>500X

95%
>30X

95%
>500X

95%
>30X

95%
>500X

95%
>30X

95%
>500X

95%
>30X

95%
>500X

95%
>30X

95%
>500X

Number of amplicons per library

Approximate number of libraries per chip

12–24

84

8

272

26

>384

48

>384

152

>384

>384

>384

>384

25–48

48

4

152

14

272

26

>384

84

>384

272

>384

>384

49–96

26

2

84

8

152

14

>384

48

>384

152

>384

243

97–192

14

1

48

4

84

8

272

26

>384

84

>384

134

193–384

8

26

2

48

4

152

14

>384

48

>384

77

385–768

4

14

1

26

2

84

8

272

26

>384

41

769–1,536

2

8

14

1

48

4

152

14

243

22

1,537–3,072

1

4

8

26

2

84

8

134

13

3,073–6,144

2

4

14

1

48

4

77

7

6,145–12,288

1

2

8

26

2

41

3

12,289–24,576

1

4

14

1

22

2

24,577–49,152

2

8

13

1

1 Discontinued, but still supported.